Second, the centrifuge rotor can be placed on the platform of the centrifuge during pre-cooling or on the test bench. However, never attempt to screw the rotor directly onto the cover. If the machine starts while the rotor is improperly secured, the cover could fly off, leading to serious accidents. Always ensure the rotor is properly positioned before starting the centrifuge.
Third, after tightening the screw cap, always check the gap between the rotor and the cover with your fingers. If there is any noticeable gap, loosen the cap and retighten it until the gap is completely eliminated. Only then should the centrifuge be started to ensure safe operation.
Fourth, during the centrifugation process, the operator must not leave the room. In case of any unusual sounds, vibrations, or malfunctions, do not turn off the power. Instead, press the STOP button immediately. Before beginning the pre-cooling process, make sure to fill out the centrifuge usage record accurately.
Fifth, never use counterfeit or low-quality centrifuge tubes. Avoid using tubes that are old, deformed, or cracked, as they may break during high-speed rotation, causing damage to the equipment and potential injury.
Sixth, you may leave the lab after completing the final routine safety check when using the centrifuge at the end of the day, especially during holidays or evening hours.
Seventh, if any malfunction occurs during the use of the instrument, contact the technical support team immediately. Do not attempt to repair or adjust internal components yourself.
For more information, visit: [http://news.chinawj.com.cn](http://news.chinawj.com.cn)
Centrifuge operator notes
When the centrifuge is in the pre-cooling phase, the cover must be securely closed. After the centrifugation process is complete, the rotor should be carefully placed on the designated test bench, and any remaining water inside the chamber should be thoroughly dried. Ensure that the centrifuge lid is left open to allow proper ventilation and cooling. It's important to note that during the pre-cooling stage, the lid must remain closed, and after use, it should be opened for safety and maintenance purposes.